← Back to Clean Air Lab Overview
Metrology Standards • 17 min read

AHAM AC-1 Standard: How Clean Air Delivery Rate (CADR) Is Empirically Measured in 1,008 Cu Ft Chambers

Peer-Reviewed by Dr. Gregory Vance, PE, CIH • Standard: ANSI/AHAM AC-1 & ISO 29463
âš¡ Top Amazon Filter Restocks & Clean Air Gear ($19 to $199)
Direct US Stock • 24-Hour Active Cookie
Replacement H13 True HEPA Filter Pack Essential Restock
Replacement H13 True HEPA Filter Pack
$29.99 US
View on Amazon
Virgin Coconut Shell Activated Carbon Pellets VOC Adsorption
Virgin Coconut Shell Activated Carbon Pellets
$24.50 US
View on Amazon
Laser Particle Counter PM2.5 / PM10 Monitor Metrology Grade
Laser Particle Counter PM2.5 / PM10 Monitor
$69.00 US
View on Amazon
AHAM Certified 300+ CADR Large Room HEPA Purifier Top Benchmark
AHAM Certified 300+ CADR Large Room HEPA Purifier
$149.99 US
View on Amazon
Electrostatic MERV 8 Washable Pre-Filter Roll Coarse Dust Saver
Electrostatic MERV 8 Washable Pre-Filter Roll
$19.95 US
View on Amazon
NDIR CO2 & MOX Metal-Oxide VOC Indoor Air Sensor Chemical Safety
NDIR CO2 & MOX Metal-Oxide VOC Indoor Air Sensor
$119.00 US
View on Amazon
Clean Air Research Sponsor

The Clean Air Delivery Rate (CADR) is the single most authoritative engineering metric for evaluating portable air cleaners. Standardized under ANSI/AHAM AC-1, it quantifies the volume of contaminant-free air delivered per minute by isolating mechanical filtration from natural particle deposition.

1. The 1,008 Cubic Foot Sealed Test Chamber

Under AHAM AC-1-2020 protocol, testing occurs inside a rigid, airtight chamber measuring exactly 10.5 ft × 12.0 ft × 8.0 ft (1,008 cu ft / 28.5 m³). The chamber is lined with electropolished stainless steel or fluoropolymer sheeting to minimize electrostatic wall losses. Temperature is regulated to 70°F ± 5°F (21°C ± 3°C) and relative humidity is locked at 40% ± 5% RH.

Challenge AerosolParticle Size Range (Geometric Mean)Aerosol Generation SourceMaximum Certified CADR Limit
Tobacco / Combustion Smoke0.09 to 1.0 microns (0.2µm mean)Research Cigarettes / Incense smolder450 CFM (765 m³/h)
Coarse Atmospheric Dust0.5 to 3.0 microns (1.0µm mean)ISO 12103-1 A2 Fine Test Dust400 CFM (680 m³/h)
Paper Mulberry Pollen5.0 to 11.0 microns (7.0µm mean)Broussonetia papyrifera (Dry dust)450 CFM (765 m³/h)

2. Differential Decay Physics (k_total - k_natural)

CADR is calculated using first-order aerosol decay differential equations. First, the test chamber is injected with challenge aerosol to reach peak concentrations (e.g., 200,000 particles/cc). Natural settling, surface adsorption, and wall deposition are tracked over 20 minutes to establish the natural decay constant (k_n):

C(t) = C_0 × e^(-k_n × t)

Next, identical aerosol levels are regenerated, and the test air cleaner is operated at maximum blower speed. The total decay constant (k_t) is logged by high-speed condensation particle counters (CPC) and optical particle sizers (OPS). The CADR is calculated by multiplying the chamber volume (V = 1,008 ft³) by the differential rate:

CADR = V × (k_total - k_natural)

Because natural settling of 10-micron pollen is much faster than 0.2-micron smoke, subtracting k_natural ensures that the CADR metric isolates only the air volume purified by the machine.

3. The AHAM Two-Thirds (2/3) Rule for Room Sizing

To achieve the standard 4.8 air changes per hour (ACH) in a room with 8-foot ceilings (an 80% particle reduction in 30 minutes), the Smoke CADR rating must equal at least two-thirds of the floor area in square feet:

Minimum Smoke CADR = Room Area (sq ft) × 0.667

For a typical 300 sq ft master bedroom, an air purifier must carry an AHAM Smoke CADR of at least 200 CFM. Running an undersized 90 CADR unit in that same space will fail to outpace ambient indoor particle infiltration from open doors and HVAC air handlers.

Laboratory Filtration Partner

Consultation & Lab Inquiries

For room air exchange modeling, hospital cleanroom CADR audits, or filter media testing: martbusinnes@gmail.com.